Bovine serum albumin mixture: averaged and individual data frames (subtracted and unsubtracted test sets)
Data type: SASBDB experimental dataStatus: PublishedCurve type: Single concentrationLast updated: 2023-06-21T07:56:39.492710+02:00
The data displayed in this entry represents the averaged 1D-SAXS profile obtained from 100 individual SAXS data frames. Included in the full entry zip-archive are each of the individual subtracted SAXS profiles as well as each individual sample and buffer frame (100 in total).
▼
1. Samples, Components and Experimental Conditions Sample & Experiment
Sample 1 · Bovine serum albumin mixture: averaged and individual data frames (subtracted and unsubtracted test sets)
2. Measurements Reported by SASBDB Reported Results
Metric
Method
Value
Uncertainty
Unit
dmax
P(r)
11.0
—
nm
i0
Guinier
11597.7
7.33
arbitrary
i0
P(r)
11640.0
—
arbitrary
mw
Experimental
72.4
—
kDa
mw
Porod
73.0
—
kDa
porod_volume
Porod
116.7
—
nm³
rg
Guinier
3.02
0.18
nm
rg
P(r)
3.067
—
nm
These values are reported by SASBDB; SAXSdb did not recalculate them from the experimental curve.
▼
3. Source Fits and Models Source Fits & Models
This entry has no source fit record.
▶
4. Source File Index Source Files
Category
Path
Download status
Size
curve
sasbdb/entries/n3/sasddn3/source/SASDDN3.dat
downloaded
87145
full_entry_zip
sasbdb/entries/n3/sasddn3/source/SASDDN3.zip
downloaded
7338376
pddf
sasbdb/entries/n3/sasddn3/source/SASDDN3.out
downloaded
117263
sascif
sasbdb/entries/n3/sasddn3/source/SASDDN3.sascif
downloaded
168798
summary
sasbdb/entries/n3/sasddn3/source/summary.json
downloaded
7412
▶
5. Experiment Description and Publication Experiment & Publication
Original Experiment Description
Synchrotron SAXS data from a solution of bovine serum albumin (mixture) in 50 mM HEPES, pH 7.5, were collected on the EMBL-P12 beam line at the PETRA III storage ring (DESY, Hamburg, Germany) using a Pilatus 2M detector at a sample-detector distance of 3.1 m and at a wavelength of λ = 0.124 nm (l(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). One solute concentration of 2.25 mg/ml was measured at 10°C. 100 successive 0.050 second frames were collected. The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.
Related Publication
Title
Machine Learning Methods for X-Ray Scattering Data Analysis from Biomacromolecular Solutions.
Authors
Franke D, Jeffries CM, Svergun DI
DOI / PMID
10.1016/j.bpj.2018.04.018 / 29874600
▼
6. Complete Source Record Complete Source Record
The following content comes directly from the SASBDB entry. Missing values are shown as “—”; SAXSdb does not infer undeclared units.
European Molecular Biology Laboratory (EMBL) - Hamburg outstation
experiment.contributor[0].affiliation[0].webpage
http://www.embl-hamburg.de/index.php
experiment.contributor[0].contributor_name
Cy M
experiment.contributor[0].contributor_surname
Jeffries
experiment.contributor[0].orcid
null
experiment.concentration_method
null
experiment.concentration_unit
null
experiment.date
2016-09-25
experiment.storage_temperature
10.0
experiment.cell_temperature
10.0
experiment.exposure_time
0.05
experiment.number_of_frames
100
experiment.wavelength
0.124
experiment.sample_detector_distance
3.1
experiment.concentration_min
null
experiment.concentration_max
2.25
experiment.sample_volume
null
experiment.flow_rate
null
experiment.s_min
0.013
experiment.s_max
3.79
experiment.total_exposure_time
null
experiment.seccolumn
null
fits
[]
estimated_volume_method
null
pddf_software
ATSAS GNOM
pddf_software_version
5.0
i0_calibration_standard
null
description
The data displayed in this entry represents the averaged 1D-SAXS profile obtained from 100 individual SAXS data frames. Included in the full entry zip-archive are each of the individual subtracted SAXS profiles as well as each individual sample and buffer frame (100 in total).
experiment_description
Synchrotron SAXS data from a solution of bovine serum albumin (mixture) in 50 mM HEPES, pH 7.5, were collected on the EMBL-P12 beam line at the PETRA III storage ring (DESY, Hamburg, Germany) using a Pilatus 2M detector at a sample-detector distance of 3.1 m and at a wavelength of λ = 0.124 nm (l(s) vs s, where s = 4πsinθ/λ, and 2θ is the scattering angle). One solute concentration of 2.25 mg/ml was measured at 10°C. 100 successive 0.050 second frames were collected. The data were normalized to the intensity of the transmitted beam and radially averaged; the scattering of the solvent-blank was subtracted.